Introduce
- IM-600 is easy-to-use illuminance meter and capable of wide range measurement.
 - It is ideal for next generation lighting to evaluate, and also there is another type of IM-600M.
 - This product can be useful for small space and installation of a equipment.
 
Feature
- Wide range measurement (0.005 – 999,000 lx)
 - Display unit and light sensing unit are Separable with optional cable(IM-600)
 - Multiple measurement function
 - Response time, screen flicker can be measured with oscilloscope
 - Zero calibration function without cap can be installed equipment
 
Application
- Development and design department for lighting equipment maker.
 - Lighting system performance evaluation for building constructor.
 - Management of light source for plant factory.
 - Maintenance and management of light in factory, school, office, etc.
 - As a measurement sensor of production line or a variety equipment.
 
Specification
| Item | Specification | |
|---|---|---|
| JIS Class | Conforms to the general AA class illuminance meter (JIS C 1609-1-2006)  | 
|
| Photo detector | Silicon photo diode | |
| Measurement Range  | 
0.005 – 999,000 lx | |
| Cosine Correction Characteristics | 【 IM-600 】 10°: Within ±1% 30°: Within ±1% 50°: Within ±5% 60°: Within ±5% 80°: Within ±10%  | 
【 IM-600M 】 10°: Within ±1% 30°: Within ±2% 50°: Within ±6% 60°: Within ±7% 80°: Within ±25%  | 
| Relative Spectral Response | 6% or less (*1) | |
| Linearity | ±2% of rdg. ±1 digit (Auto Range) | |
| Measurement Mode  | 
Auto / Manual (6 range) | |
| Measurement Function  | 
Standard measurement: Illuminance(lx), Integrated illuminance(lx・h), Integration time(h), Luminous intensity(cd) Deviation measurement: Illuminance difference(lx), Illuminance ratio(%)  | 
|
| Calibration Function  | 
CCF (Color Correction Factor) setting | |
| Analog Output | 0 – 3V | |
| Interface | USB | |
| Power Source | Size AA dry-cell battery×2 or Dedicated adapter (Option) | |
| Use conditions | Temperature : -10 – +40℃, Humidity : 85% RH or less (No condensation)  | 
|
| Outer dimensions (W×D×H)  | 
188×65×34mm, Sensor Unit (IM-600M): 16×21mm | |
| Weight | IM-600: 200g, IM-600M: 230g *Non-dry cell battery | |
Accessories